Low-level gas multicounter for C-14 dating of small samples: Electronic, numerical and shielding optimisation

Authors

  • L. Kaihola
  • H. Kojola
  • H. Polach
  • P. Mäntynen
  • J. Tervahauta
  • E. Soini

Abstract

Up to 14 methane samples can be dated simultaneously in our compact gas multicounter. Sample detectors are 10 ml (NTP) in volume each. They are made of copper and linked to form two 7 detector rigid assemblies which are filled in situ. Monitoring of the counting conditions is enabled through multichannel analysis of the cosmic pulse height spectrum, which shows the changes in gas amplification due to impurities or leakage. HV is set (and adjusted) automatically using the cosmic peak. All individual events are stored on disc, including pulse height (PH), risetime (RT) (both 256 Ch), time of arrival (TA), detector identification, anticoincidence status and elapsed and live time. Software programs analyse and validate data. Numerical discrimination and manipulations of counting parameters can be performed without destroying the original data set. Statistical quality control is based on chi-square and Poisson distribution of count rates around their mean in user defined energy regions as weil as time of arrival of pulses mode. TA analysis offers the user an early means for recognizing some types of system malfunction that otherwise might remain undetected for Jong periods of time. RT analysis is used to discriminate sample beta pulses from environmental radiation pulses, resulting in a low background with compact and relat ively inexpensive shielding. The automatic high voltage setting, PH, RT and TA electronics as weil as the liquid scintillation anticoincidence systems are applicable to all existing gas counting systems.

Delivery of the gas multicounter to the Australian National University is to take place at the end of the year 1984.

Downloads

Published

1985-03-25